Due to the special design of the standard cross-section in each roller position, LOGICA roller clamps are especially suitable for use in the enteral feeding therapy. Fibre-filled solutions will not clog the tube's cross-section within common drop rates.
As a result of the design features of LOGICA's roller clamps, an adjusted drop rate will be stable within small tolerances. The regulation span is not influenced by the movement direction of the roller; the control behaviour remains the same. The easy and fast adjustment of the roller and the large regulation span allows you to set the desired drop rate fast and effortlessly. As the control behaviour barely changes across the entire regulation span, precise adjustment of the desired drop rate is possible for every infusion solution, independent of the viscosity.
The LOGICA Roller Clamp RC2OOO MULTI can be used in enteral and parenteral infusion sets. The Roller Clamp RC2OOO MULTI is available for tubing dimensions of 3.0 x 4.1 mm and 3.1 x 4.3 mm.
- Enhanced development of the proven LOGICA RC2OOO
- particulary suitable for the use in enteral feeding sets and in infusion and transfusion sets
- optimal regulating design for easy adjustment of drip rate
- low deviation of adjusted drip rate (±10%)
- good reproducibility of desired adjustment
- Tube's cross-section within common drop rates
Adjusting length | ≥ 9 mm (between 10 and 150 drips/min) |
---|---|
Drip rate deviation after 2 h | ≤ 10 % (between 10 and 150 drips/min) |
Volumetric flow rate at end of adjustable stretch | ≥ 500 drips/min (1500 ml/h) |
In closed position | air tight at 0.5 bar for min. 5 minutes |
In fully opened position | tube's cross-section completely opened |
Sterilization compatibility | ETO-Gas, Electron Beam Irradiation and Gamma Irradiation up to 50 kGy (cumulative) |
Adjusting length | ≥ 9 mm (between 10 and 150 drips/min) |
---|---|
Drip rate deviation after 2 h | ≤ 10 % (between 10 and 150 drips/min) |
Volumetric flow rate at end of adjustable stretch | ≥ 150 drips/min (450 ml/h) |
In closed position | air tight at 0.5 bar for min. 5 minutes |
In fully opened position | tube's cross-section completely opened |
Sterilization compatibility | ETO-Gas, Electron Beam Irradiation and Gamma Irradiation up to 50 kGy (cumulative) |
The LOGICA Roller Clamp RC2OOO 4,0x5,4 has been developed for use in enteral infusion therapy.
- optimal regulating design for easy adjustment of drip rate
- low deviation of adjusted drip rate (±10%)
- good reproducibility of desired adjustment
- particulary suitable for the use in enteral feeding sets
- tubing clamp for save fastening of the tubing
- easy automatic or hand operated assembly onto the tubing
Adjusting length | ≥ 8 mm (between 10 and 150 drips/min) |
---|---|
Drip rate deviation after 2 h | ≤ 10 % (between 10 and 150 drips/min) |
Volumetric flow rate at end of adjustable stretch | ≥ 250 drips/min (750 ml/h) |
In closed position | air tight at 0.5 bar for min. 5 minutes |
In fully opened position | tube's cross-section completely opened |
Sterilization compatibility | ETO-Gas |
for enquiries and services