The LOGICA Lid Valve is designed for installation in housings provided by the customer. It can be used as an air valve, e.g. in drip chambers, or as a check valve in infusion sets.
With its proven valve mechanism, the lid valve ensures a low opening pressure and fast closure in the case of back pressure. The safe function of the valve is guaranteed, independently of its position.
- silicon valve for assembly into existing housing
- usable as air valve or check valve in infusions sets
- low opening pressure
- material is latexfree and DEHP-free
Valve Specification “Lid Valve”
Opening pressure when first used (dry) | 5 mbar |
---|---|
Opening pressure when first used (wet) | 20 mbar |
Counterflow pressure resistance | 1 bar |
Blocking performance (leak rate) | 0.003 ml/min |
Position sensitivity | independent of position |
Stand-by position | open |
Sterilization compatibility | ETO-Gas, Electron Beam Irradiation and Gamma Irradiation up to 50 kGy (cumulative) |