Check valve V2OOON IV with “T-connector”
The LOGICA Check Valve V2OOON with a “T-connector” is a new supplement to the well-known V2OOO series. It has been developed for applications where it is required to connect two parallel infusions lines into one valve-protected line. The male Luer Lock connector consist of a rotatable lock ring. All Luer Lock connectors are according to ISO 80369-7. The membrane valve has the same outstanding properties which are known from the V2OOO Check Valve series.
- check valve closed in stand-by position
- function independent from position of use
- materials are latexfree and DEHP-free
The new V2OOON Check Valve is available in two different specifications.
Valve specification „V2OOON“
| V2OOO V Specification (high opening pressure) | V2OOO IV Specification (low opening pressure) |
---|---|---|
Opening pressure (when first used) | 200 ± 50 mbar | ≤ 6 mbar |
Opening pressure (after closure of the check valve) | 200 ± 50 mbar | ≤ 12 mbar |
Counterflow pressure resistance | 21 bar | 5 bar |
Pressure resistance in flow direction | 21 bar | 6 bar |
Volumetric flow rate | ≥ 170 ml/min at a pressure of 500 mbar | ≥ 150 ml/min |
Blocking performance (leak rate) | ≤ 0,1 ml/h | ≤ 0,1 ml/h |
Position sensitivity | independent of position | independent of position |
Stand-by position | closed | closed |
Sterilization compatibility | ETO-Gas, Electron Beam Irradiation and Gamma Irradiation up to 50 kGy (cumulative) | ETO-Gas, Electron Beam Irradiation and Gamma Irradiation up to 50 kGy (cumulative) |