Check valve V2OOO High Flow
The LOGICA Check Valve V2OOO High-Flow has been especially developed for applications where a high inflow of fluids is required in connection with a low flow resistance.
The valve is closed in stand-by position. It opens in flow direction at low pressures (≤20 mbar). The safe function of the valve is guaranteed independently of its position. The connections of the high-flow check valve are designed for tubes with an outer diameter of 8.8 mm or 6.8 mm.
- check valve closed in stand-by position
- function independent from position of use
Valve Specification „V2OOO High Flow“
Opening pressure (when first used) | ≤ 20 mbar |
---|---|
Opening pressure (after closure of the check valve) | ≤ 20 mbar |
Counterflow pressure resistance | 1 bar |
Pressure resistance in flow direction | 1 bar |
Volumetric flow rate | ≥ 1000 ml/min at a pressure of 100 mbar ≥ 3000 ml/min at a pressure of 1 bar |
Blocking performance (leak rate) | ≤ 0.5 ml/h |
Position sensitivity | independent of position |
Stand-by position | closed |
Sterilization compatibility | ETO-Gas, Electron Beam Irradiation and Gamma Irradiation up to 50 kGy (cumulative) |