Air valve V2OOO
The LOGICA Air Valve V2OOO with integrated bacterial filter is usable in infusion and transfusion sets and in sets for enteral feeding therapy.
The seal mechanism secures a safe function of the valve, independent of the valve's position. Pressure loads against the flow direction also have no influence on the function of the valve. The membrane valve is closed in stand-by position. Reflow to close the valve is not required. The bacterial filter secures the safe retention of bacteria. Wetting of the bacterial filter is effectively prevented by the valve.
- usable for example in spikes of drip chambers
- valve closed in stand-by position
- function independent from position of use
- integrated bacterial filter
- high air flow
- materials are latexfree and DEHP-free
Valve Specification „Air Valve V2OOO“
Opening pressure (when first used) | ≤ 8 mbar |
---|---|
Counterflow pressure resistance | 2 bar |
Volumetric flow rate for an infusion set | Enables a volumetric flow rate of ≥ 100 ml/min through an infusion set according to ISO 8536-4 |
Position sensitivity | independent of position |
Stand-by position | closed |
Hydrophobicity of the filter | 140 cm H2O |
Filter separation efficiency | > 99.985 % (Filter efficiency is measured in accordance with EN 143 at 30 l/min.) |
Sterilization compatibility | ETO-Gas, Electron Beam Irradiation and Gamma Irradiation up to 50 kGy (cumulative) |