_Check Valve LOGICA V2OOOrueckschlagventil v2OOO

 

 

_Product Properties:

 

_check valve closed in stand-by position
_function independent from position of use
_low priming volume
_materials are latexfree and DEHP-free

 

 

_Technical Specification:

 

_Opening pressure ≤ 12 mbar (also after counterflow pressure)
_Counterflow pressure resistance 6 bar
_Burst pressure 6 bar
_Blocking performance leakage rate ≤ 0.1 ml/h
(meets the requirements of DIN 58362-7 resp. ISO 8536-12)
≥ 130 ml/min
_Volumetric flow rate*  ≥ 130 ml/min
_Sterilization ETO-gas, γ-irradiation up to 50 kGy (cumulative)

 

*: Test according to ISO 8536-12. Tested with infusion sets according to ISO 8536-4.
Test medium 0.9 % saline solution.
Static pressure 100 mbar (1 mH2O).

 

pdf Check Valve V2000